This model succeeded LAS-3000mini which was equipped with Fujifilm's proprietary high sensitive and high resolution CCD technology and further evolved to the system with various capable applications with wider line-ups of optional product, such as epi-blue illuminator, epi-white illuminator or UV trans-illuminator. In addition to all-in-one compact body, quick focusing by control from a computer and quick response by USB 2.0 interface are available.
Fujifilm has discontinued sales of this product. An equivalent product is available through GE Healthcare.
By rotating pixels 45 degrees to form an interwoven layout, the Super CCD's pixel pitch in the horizontal and vertical directions is narrower than in the diagonal direction, achieving higher horizontal and vertical resolution.
![[Photo] Super CCD chip](img/index/pic_01.jpg)
Super CCD chip
![[Figure] Octagonal pixel-interwoven layout](img/index/pic_02.jpg)
Octagonal pixel-interwoven layout
The analyzer incorporates a FUJINON, a strikingly bright lens with an F-number of 0.85. This lens has been especially designed to make full use of the advantages of Fujifilm's proprietary Super CCD chip, and is excellent for capturing images from distances as short as several tens of centimeters.
In its design, optical expertise developed through professional applications such as broadcasting TV camera lens is fully exploited.
![[Photo] FUJINON LENS VRF43LMD 3 F0.85](img/index/pic_03.jpg)
FUJINON LENS VRF43LMD 3 F0.85
The same grade of camera and lens as in LAS-4000, the highest-grade model of the LAS series, are used. A compact body is made possible while still retaining the high resolution and high performance of LAS-4000. Free from the necessity for dark rooms and chemicals, the analyzer can be placed on the top or side of an ordinary desk, and requires no special expertise for operation.
The self-explanatory and user-friendly software allows specification of all image capture parameters, including sensitivity, resolution and image methods. The software automatically carries out all image correction functions. In increment mode, the signal increase is shown in real time, by capturing and automatically displaying up to 16 images. With this software users can create their own settings and methods. A photograph of the sample can be made and overlaid with the signal image.
Method/Tray position Screen
Exposure Increment Screen
Includes; UV 2020 transilluminator, 605DF40 optical filer for EtBr detection, UV-Dia tray, Gel sheet, etc.
Includes; Housing box for UV transilluminator, Filter frame, Filter housing, Bottom shield
Includes; EPI Blue / White Illuminator, Y515 optical filter, etc.
A non-parallax (NP) tray is used to eliminate the parallax effect when imaging chemiluminescence in the 96-well plate. Up to two plates can be placed.
![[Photo] NP Tray](img/index/pic_06.jpg)
| CCD chip | Fujifilm Super CCD Area Type chip |
|---|---|
| Number of pixels | 3.2 million pixels |
| Pixel size | Approx. 11 µm |
| Cooling | Two-stage thermoelectric module with air circulation |
| Cooling temperature | Down to -30°C (When room temperature is below 28°C.) |
| Dynamic range | Four orders of magnitude |
| Focusing and diaphragm | Automatic, remote operation |
| Gradation | 16 bits |
| Exposure mode | Automatic / manual (normal / incremental / repetitive) |
| Exposure time | Automatic / manual (1 / 100 seconds to 30 hours) |
| Pixel correction | Dark-frame correction, flat-frame correction, distortion correction, etc. |
| Image quality correction | Binning and smoothing |
| Image size | Up to 12 MB (formats : FUJI and TIFF) |
| Read pixel size | Down to 30 µm |
| Maximum sample size | 18 x 12 cm |
| Interface | USB2.0 |
| High-sensitivity lens | FUJINON VRF43LMD3 F0.85 |
|---|
| Line frequency | 50 / 60 Hz |
|---|---|
| Temperature | 15 - 28℃ |
| Humidity | 30 - 70% (no condensation) |
| Supply Voltage | 100 - 240V |
| Power Consumption | Approx. 0.3 kVA |
| Camera head and dark box | 440 (W) x 670 (H) x 370 (D) mm 440 (W) x 750 (H) x 380 (D) mm / With UV transilluminator |
|---|
| UV 2020 Transilluminator set | Includes; UV 2020 transilluminator, 605DF40 optical filter for EtBr detection, UV-Dia tray, Gel sheet |
|---|---|
| UV-Bay set | Includes; Housing box for UV transilluminator, Filter frame, Filter housing, Bottom shield |
| Epi-Blue Illuminator set | Includes; EPI Blue / White Illuminator, Y515 optical filter |
| Non-parallax tray | |
| Optical Filter |
|
| OS |
|
|---|
| Chemiluminescence / Bioluminescence | ECL™, ECL Plus™, ECL Advance™, Lumi-Light Plus, SuperSignal®, CDP-Star®, CSPD®, Renaissance, Bright-Star™, Luciferase |
|---|---|
| Fluorescence / Blue epi-illumination* | SYBR® Green I, SYBR® Green II, SYBR® Gold, SYPRO® Ruby, SYPRO® Orange, SYPRO® Tangerine, FITC, FAM™, AttoPhos™ |
| Fluorescence / trans-UV** | Ethidium Bromide |
| Digitizing / White epi-illumination*** | NBT / BCIP, Silver Staining, CBB |
Gel documentation is available by digitizing