Classification Parameter Measurement range *1 Measurement time
(min.)
Unit (A) Unit (B)
Biochemical tests Enzymes ALP 14 – 1183 U/L 0.23 – 19.76 μ Kat/L 4
AMYL 10 – 1800 U/L 0.17 – 30.06 μ Kat/L 5
CHE 5 – 500 U/L 0.08 – 8.35 μ Kat/L 4.5
CKMB 1 – 300 U/L 0.02 – 5.01 μ Kat/L 5
CPK 10 – 2000 U/L 0.17 – 33.40 μ Kat/L 4
GGT 10 – 1200 U/L 0.17 – 20.04 μ Kat/L 5
GOT/AST 10 – 1000 U/L 0.17 – 16.70 μ Kat/L 4
GPT/ALT 10 – 1000 U/L 0.17 – 16.70 μ Kat/L 4
LAP 10 – 500 U/L 0.17 – 8.35 μ Kat/L 4
LDH 50 – 900 U/L 0.84 – 15.03 μ Kat/L 2
LIP 20 – 1000 U/L 0.33 – 16.70 μ Kat/L 5
General chemistry ALB 1.0 – 6.0 g/dL 10 – 60 g/L 6
BUN 5.0 – 140.0 mg/dL 1.79 – 49.98 mmol/L 4
Ca 4.0 – 16.0 mg/dL 1.00 – 4.00 mmol/L 4
CRE 0.2 – 24.0 mg/dL 18 – 2122 μmol/L 5
DBIL 0.1 – 16.0 mg/dL 2 – 274 μmol/L 5
GLU 10 – 600 mg/dL 0.6 – 33.3 mmol/L 6
HDL-C 10 – 110 mg/dL 0.26 – 2.84 mmol/L 6
IP 0.5 – 15.0 mg/dL 0.16 – 4.84 mmol/L 5
Mg 0.2 – 7.0 mg/dL 0.08 – 2.88 mmol/L 4.5
NH3 10 – 500 μg /dL 7 – 357 μmol/L 2
TBIL 0.2 – 30.0 mg/dL 3 – 513 μmol/L 6
TCHO 50 – 450 mg/dL 1.29 – 11.64 mmol/L 6
TCO2 5 – 40 mmol/L 5 – 40 mmol/L 5
TG 10 – 500 mg/dL 0.11 – 5.65 mmol/L 4
TP 2.0 – 11.0 g/dL 20 – 110 g/L 6
UA 0.5 – 18.0 mg/dL 30 – 1071 μmol/L 4
Electrolytes Na 75 – 250 mEq/L 75 – 250 mmol/L 1
K 1.0 – 14.0 mEq/L 1.0 – 14.0 mmol/L
Cl 50 – 175 mEq/L 50 – 175 mmol/L
Immunological test CRP 0.3 – 7.0 mg/dL 3 – 70 mg/L 5
Comprehensive S-panel TP, ALB, ALP, GLU, TBIL, IP, TCHO, GGT, GPT/ALT, Ca, CRE, BUN
Small Screening S-panel ALP, TBIL, GGT, GPT/ALT, CRE, GOT/AST
  • * There are parameters which may not be available in your area. For details please contact your local distributor.
  • *1 Unit (A) or (B) is available
Calculations

Calculated Parameter

Indication

Unit

Equation

LDL Cholesterol

LDL

mg/dL

LDL-C = TCHO value - (HDL-C vlaue + TG value/5)

mmol/L

LDL-C = TCHO value - (HDL-C value + TG value/2.2)

non-HDL Cholesterol

non-HDL

mg/dL or mmol/L

non-HDL = TCHO value - HDL-C value

Globulin

GLOB

g/dL or g/L

GLOB = TP value - ALB value

Albumin/Globulin ratio

ALB/GLOB

-

ALB/GLOB = ALB value / (TP value - ALB value)

BUN/Creatinine ratio

BUN/CRE

-

BUN/CRE = BUN value / CRE value

GOT/GPT ratio (AST/ALT ratio)

GOT/GPT (AST/ALT)

-

GOT/GPT=GOT value / GPT value (AST/ALT=AST value / ALT value)

Sodium/Potassium ratio

Na/K

-

Na/K=Na value / K value

Anion Gap

Anion Gap

mEq/L or mmol/L

Anion Gap = Na value - (Cl value + TCO2 value)

 

Main specications
Measurement test

Colorimetry 28 tests, Electrolytes 3 tests

Throughput

Colorimetry 180 test/hour, Combined 190 test/hour

Number of sample rack

5

Number of incubator cell

Colorimetry 13, Electrolytes 1

Measurement time

Colorimetry 2 to 6 minutes/test,

Electrolytes 1 minute/3 tests (Na-K-Cl)

Sample type

Plasma, Serum, Whole blood*2

Sample volume

Colorimetry 10μL/test,

Electrolytes 50μL/3 tests (Na-K-Cl), CRP 5μL/test

Data transmission to PC

RS 232C (1 port), USB (1 port), LAN (1 port)

Data print

Thermal Printer

Electrical requirements

Single phase AC; 100 - 240 V ±10%; 50 to 60 Hz

Display

7-inch color touch panel

Dimensions

500 (W) ✕ 380 (D) ✕ 410 (H) mm

Weight

Approx. 33kg

Operating temperature

15 to 32˚C (59 to 89F)

Operating humidity

30 to 80%RH

  • *2 NH3-W: Whole blood only
  •     NH3-P: Plasma only
  •     Na-K-Cl: Plasma, Serum, Whole blood
  •     Other test items: Plasma, Serum
DRI-CHEM NX700 Series
  NX700 NX700i
Electrolyte tests
Plasma Filter Function -
Automatic dilution
  • *Please contact your local distributor for availability.
Option Items
Sample Rack

For φ16 × 100 mm blood collection tube
For φ13 × 100 mm blood collection tube
For φ13 × 75 mm blood collection tube
For 1.5 mL Fuji tube 
For 0.5 mL Fuji tube
For φ16 × 100 mm blood collection tube (when using PF)
For φ13 × 100 mm blood collection tube (when using PF)
For φ13 × 75 mm blood collection tube (when using PF)