The content on this page is intended to healthcare professionals and equivalents.
At the end of the course, you will be able to :
- Implement specific mammography techniques
- Perform appropriate internal quality controls
Have already used a mammography unit for the first time.
Technicians, Doctors / Radiologists, Physicists, Biomedical Engineers
3 days (9:00 am - 17:30 pm)
In our technical center or on your facilities
Please contact us
- PowerPoint slideshow
- Practical application to real or simulated situations
- Specific equipment: mammograph, MTM100 phantom, plexiglass plate
- Training supervision: provided by the training manager
- Instructor: Application engineer, trainer and employee of FUJIFILM Healthcare France SAS
- Initial assessment: test or role-play to assess each trainee's skills and weaknesses.
- Ongoing assessment of progress (during application exercises): checks on acquisition of the concepts studied and ability to apply them, identification of areas for improvement.
- Knowledge test at the end of the course.
- End-of-training certificate issued.
- Theoretical and methodological contributions, with real-life examples illustrating the application of the concepts covered.
- Individual application of the training with case studies of mammography examinations
- Case studies, role-playing scenarios: interactive analysis, identification of good/bad approaches and corrections.
9h00 - 10h30
- Round-table discussion of practices and difficulties encountered
- Participants' expectations
- 1st survey : identification of training needs
10h30 - 12h30
- Function and purpose of the various components of a mammography unit
- Analysis and answers to the questionnaire
- X-ray production
- Mammography at Fujifilm
14h30 - 17h00
- Knowledge of the mammography machine's general components
- Functioning and benefits of the various components of a mammography unit
Incidences in mammography - Operating principles of tomosynthesis and 2D synthetics
- Lesions and screening
(use of a breast phantom)
09h00 - 12h30
- Switching on & off and startup the machine
- 2D image acquisition
- Magnification kit installation
- Dose reduction and image quality enhancement tools
- Tomosynthesis (slice acquisition and scrolling)
- Image processing tools
- Image and patient file management tools
14h00 - 17h30
- Presentation of daily quality control (AAPM TG 18 test pattern and use of luxmeter)
- Presentation of weekly quality control: MTM100, 2D and 3D homogeneity on Plexiglas plate
- Presentation of monthly quality control: mammograph integrity
- Cleaning and disinfection of various elements
- Awareness of energy consumption
- Awareness of the challenges of personal data confidentiality (RGPD).
09h00 - 12h30
- 2D routine screening examinations (with and without prostheses)
- Specific examinations: enlarged, localized
14h00 - 16h00
- Tomosynthesis exams
16h00 - 17h30
- Survey to validate skills
- Analysis of practices, conclusions and ideas
- Improvement actions
Quotation on request at : formations_hcfr@fujifilm.com