In order to provide you with a better browsing experience and to improve our site functionality, we use cookies and other tracking technologies. Detailed information on the use of cookies on our site and how to opt out is provided in our Cookie Policy. By clicking into any content on this site, you consent that we can store and access cookies and other tracking technologies as described in our Cookie Policy.

Philippines

FUJIFILM COVID-19 Ag Test

Exclusive reagent for COVID-19 Ag(Target specimen: Nasopharyngeal swab)

The content on this page is intended to healthcare professionals and equivalents.

Product Features

  • Uses Silver Amplification technology for high detection sensitivity
  • Time to results: Approximately 15 minutes
  • Compatible with both visual and reader detection
  • Nasopharyngeal swab

Silver Amplification technology

The immunochromatographic assay was designed based on silver halide photography to amplify the size of gold colloidal particles approximately a 100-fold, leading to the improvement of detection sensitivity. This offers a greater ability to detect even in small amounts of antigens.

How to use

  • * Refer to the Instruction for Use for official procedures.

Performance results

Clinical performance was established using retrospectively collected 225 samples (nasopharyngeal swabs).

Sensitivity: 90.9% (110/121, 95% CI: 84.3-95.4%)
Specificity: 100.0% (104/104, 95% CI: 96.5-100.0%)

  RT-PCR
Positive Negative Total
FUJIFILM COVID-19 Ag Test Positive 110 0 110
Negative 11 104 115
Total 121 104 225

Sensitivity according to the RT-PCR Ct value

  RT-PCR Ct value
< 26 < 28 < 30 < 35
FUJIFILM
COVID-19
Ag Test
Positive 100 107 110 110
Negative 0 3 6 11
Total 100 110 116 121
Sensitivity 100.0% 97.3% 94.8% 90.9%
95% CI 96.4-100.0% 92.2-99.4% 89.1-98.1% 84.3-95.4%

Kit contents

FUJIFILM COVID-19 Ag Test (10 tests)

  • Test Cartridge × 10 tests
  • Extraction Reagent Solution R1 × 10 × 0.4mL
  • Sterile Swab × 10 pieces
  • Tube Rack (for Extraction Reagent Solution R1) × 1 piece
  • Filter (for Extraction Reagent Solution R1) × 10 pieces
  • Package Insert × 1 piece

Additional instrument (Optional, not included)

  • FUJIFILM SILVAMP Reader

left : FUJIFILM COVID-19 Ag Test
right : FUJIFILM SILVAMP Reader